Contract award record
ITT_121341 - Cryo-free QHR Fridge
Notice details
- Published
- Category
- goods
- Buyer region
- London
- Source
- View official notice
Public buyer and contract winner
Award value£159,600
- AwardedCompleted
- StartCompleted
- EndCompleted
Work description
The National Physical Laboratory (NPL) requirement is for a refrigerator to house one of our Cryogenic Current Comparators (CCC). Currently the system is mounted in a dewar of helium to provide the necessary temperature, but this requires refilling periodically and we wish to move to a closed setup which does not need additional cryogens for sustained periods of operation.
The asset must therefore:
1) Allow the existing CCC components to be mounting in the cold space of the fridge. 2) Hold a steady temperature of 4 K or less for a sustained period. 3) Not require any additional liquid cryogens for cool-down or whilst running.